Tissue-to-air ratio
From Infogalactic: the planetary knowledge core
Lua error in package.lua at line 80: module 'strict' not found.
Tissue-to-air ratio (TAR) is a term used in radiotherapy treatment planning to help calculate absorbed dose to water in conditions other than those directly measured.
Definition
The TAR at a point in a water phantom irradiated by a photon beam is taken to be the ratio of the total absorbed dose at that point to the absorbed dose at the same point in a minimal-scatter phantom with just-sufficient build-up.[1][2]
Tissue-air ratio is defined as the ratio of the dose to water at a given depth to the dose in air measured with a buildup cap:
[TEX] TAR=frac{D(f,z)}\{D(f,0)} [/TEX]
where D(f,z) is the dose at a given depth z and distance focus-detector f; and D(f,0) is the dose in air (z=0).
Measurements for each are taken using an ion chamber for identical source to detector distances and field sizes.[3]