Tissue-to-air ratio

From Infogalactic: the planetary knowledge core
Jump to: navigation, search

Lua error in package.lua at line 80: module 'strict' not found.

Tissue-to-air ratio (TAR) is a term used in radiotherapy treatment planning to help calculate absorbed dose to water in conditions other than those directly measured.

Definition

The TAR at a point in a water phantom irradiated by a photon beam is taken to be the ratio of the total absorbed dose at that point to the absorbed dose at the same point in a minimal-scatter phantom with just-sufficient build-up.[1][2]

Tissue-air ratio is defined as the ratio of the dose to water at a given depth to the dose in air measured with a buildup cap:

[TEX] TAR=frac{D(f,z)}\{D(f,0)} [/TEX]

where D(f,z) is the dose at a given depth z and distance focus-detector f; and D(f,0) is the dose in air (z=0).

Measurements for each are taken using an ion chamber for identical source to detector distances and field sizes.[3]

See also

References

  1. Johns H. E. and Cunningham J. R. (1983). The Physics of Radiology. Charles C. Thomas Publ.
  2. Hendee W., Ibbott G. and Hendee E. (2005). Radiation Therapy Physics. Wiley-Liss Publ. ISBN 0-471-39493-9.
  3. Faiz M. Khan. (2010) "The Physics of Radiation Therapy " Lippencott, Wilkins and Williams Publ.